Home › CARDIOQUIP, LLC › CardioQuip Modular Cooler-Heater (MCH)
CardioQuip Modular Cooler-Heater (MCH)
CARDIOQUIP, LLC · Model MCH-1000(i) · UDI-DI 00860000846103
DWC
Prescription
1
510(k) clearances
9,255
Adverse events (204 deaths)
52
Recalls
10
Facilities
Description
The CardioQuip Modular Cooler-Heater (MCH) is indicated to supply temperature-controlled water to heat exchange devices (e.g.; CPB Heat Exchangers, Cardioplegia Heat Exchangers, and thermal regulating system blankets) to help control a patient's temperature during cardiopulmonary bypass procedures lasting not longer than six hours.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2023-10-24 | CardioQuip, LLC | There is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection. |
| 2023-10-24 | CardioQuip, LLC | There is the potential that cooler-heater devices may contain bacterial contamination which could result in patient infection. |
| 2023-10-19 | CardioQuip, LLC | There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components. |
| 2023-10-19 | CardioQuip, LLC | There is a potential for the heating-chamber to malfunction resulting in melting/burning of the heater chamber components. |
| 2023-08-03 | LivaNova Deutschland GmbH | The firm's labeling (Quick Reference Guide) contains incorrect information in the maintenance schedule for the Heater-Cooler device. |
| 2021-07-14 | Maquet Medical Systems USA | There is the potential for contaminated water to enter parts of the Heater-Cooler Unit circuits. There is a potential risk of bacterial contamination including Mycobacteria species |
| 2021-06-24 | CardioQuip, LLC | There is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potenti |
| 2021-06-17 | Gentherm Medical, LLC | There is a potential risk of device contamination and patient infection associated with the device because of the potential for organisms (including Nontuberculous mycobacteria NTM |
| 2021-05-12 | CardioQuip, LLC | Labeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices. |
| 2021-05-12 | CardioQuip, LLC | Labeling does not include guidance for proper water-quality maintenance and device inspection for heater cooler devices. |
Query this across every FDA database — by API.
Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.
Get your free API keyRead the docsData compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.