FDA Device Intelligence
HomeInspire Medical Systems, Inc. › Inspire

Inspire

Inspire Medical Systems, Inc. · Model 900-008-051 · UDI-DI 00855728005253

MNQ Prescription
0
510(k) clearances
3,486
Adverse events (4 deaths)
6
Recalls
15
Facilities

Description

The sleep remote is an external device that is about the size of a cell phone and is used by the patient to activate the Inspire system before they go to sleep. The patient places the remote over the implanted IPG site, and uses push buttons on top of the programmer to: a.) turn the therapy ON or OFF, b.) temporarily suspend therapy, or c.) make adjustments to the stimulation strength (within limits set by the physician). The patient’s sleep remote also allows patients to check the status of the IPG battery. There are both visual and audio indicators that provide information on the IPG status (therapy ON/OFF, therapy paused, stimulation level and battery status). The Inspire sleep remote will only communicate with the Inspire IPG.

Recalls

DateFirmReason
2026-05-11Inspire Medical Systems Inc.Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By D
2026-05-11Inspire Medical Systems Inc.Products are labeled with incorrect Use By Date, extending beyond the validated three-year shelf life from date of manufacturer. Material degradation of a product past its Use By D
2024-07-31Inspire Medical Systems Inc.Labeling error with a mismatch between the shelf box serial number label and the serial number of the lead provided in the box.
2024-07-31Inspire Medical Systems Inc.Labeling error with a mismatch between the shelf box serial number label and the serial number of the lead provided in the box.
2024-05-16Inspire Medical Systems Inc.There is a potential manufacturing defect which can lead to electrical leakage in the sensing circuit which causes a system malfunction and requires a revision surgery to replace t
2018-06-12Inspire Medical Systems Inc.Incorrect use-by date on the device registration/patient file labels.