REAADS Monoclonal Free Protein S Antigen
CORGENIX MEDICAL CORPORATION · Model 051-001 · UDI-DI 00855360006014
Description
INTENDED USE An enzyme-linked immunosorbent assay (ELISA) for the quantitative determination of Free Protein S Antigen in citrated human plasma. For In Vitro Diagnostic Use. SUMMARY AND EXPLANATION OF THE PROTEIN S TEST Protein S is a vitamin K-dependent protein synthesized in the liver, vascular endothelium, and megakaryocytes, which plays an important physiologic role in the Protein C Anticoagulant System. This anticoagulant system is one of the major regulators of hemostasis by inhibiting clot formation and by promoting fibrinolysis. Protein S functions as a cofactor for activated Protein C on the vascular membrane to facilitate the degradation of clotting factors Va and VIIIa, down-regulating clot formation. In normal plasma approximately 40% of Protein S circulates as a free molecule, while 60% is complexed with C4b, a plasma protein of the classical complement pathway. Only Free Protein S is functionally active and able to bind to activated Protein C, while the complexed form of Protein S is not. Protein S deficiency, either congenital or acquired, may lead to serious thrombotic events such as thrombophlebitis, deep vein thrombosis, or pulmonary embolism. The prevalence of Protein S deficiency has been estimated to be less than 1 case per 300 in the general population. Two-thirds of patients with a congenital deficiency of Protein S (levels less than 50% of normal) may present with venous thrombosis in young adulthood. In young patients (<35 years) with a history of thrombosis, the prevalence may be as high as 15 to 18%. Acquired Protein S deficiency may be seen during pregnancy, oral contraceptive or oral anticoagulant therapy, liver disease, diabetes mellitus, postoperative complications, septicemia, and various inflammatory syndromes. A decreased Protein S activity in plasma may be the result of low concentrations or abnormal function of the Protein S molecule. Refer to Product Package Insert.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2026-07-29 | Diagnostica Stago, Inc. | A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified |
| 2026-02-04 | Diagnostica Stago, Inc. | The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values. |
| 2023-02-01 | Aniara Diagnostica LLC | New warnings and precautionary statements on product labelling: Reagent 2 (R2) of BIOPHEN Protein C 5 contains the chemical substance - Cesium Chloride which changed classification |
| 2020-03-04 | Diagnostica Stago, Inc. | A risk of sample-to-sample cross-contamination of Emicizumab for factor VIII assays on Stago instruments, clotting times will be shortened that may cause factor VIII levels report |
| 2017-04-04 | Helena Laboratories, Inc. | Internal components of the Protein S ELISA Antigen Kit (Lot: 3-16-5292) were mislabeled. The Protein S Conjugate Solution is labeled as Substrate, and the Substrate is labeled as |
| 2015-12-18 | Instrumentation Laboratory Co. | This Lot is not meeting labeled on-board instrument stability claims of 5 days. |
| 2015-11-19 | Siemens Healthcare Diagnostics, Inc. | Factor VIII Chromogenic Assay Lot 44637 is showing an increased occurrence of irregular curves when used for measurement on Sysmex CA-1500, CA-7000, CS-2000i, CS-2100i, CS-5100 Sys |
| 2012-10-24 | Sekisui Diagnostics Llc | Spectrolyse PAI-1 activity assay, producing lower than expected absorbance values 0 IU/ml standard reducing the slope of the standard curve requiring more frequent sample dilutio |
| 2012-08-28 | Sekisui Diagnostics Llc | SPECTROLYSE PAI-1 activity assay producing lower than expected absorbance values with "0" standard. This will reduce the slope of the standard curve requiring the user to make mor |
| 2012-04-11 | Instrumentation Laboratory Co. | Falsely elevated levels may be reported with the HemosIL von Willebrand Factor Activity assay. |