FDA Device Intelligence
Home › LITECURE LLC › LightForce PRO

LightForce PRO

LITECURE LLC · Model LTC-900 · UDI-DI 00854103007004

ILY Prescription
10
510(k) clearances
294
Adverse events (4 deaths)
20
Recalls
498
Facilities

Description

Mains powered, infrared phototherapy device emitting up to 9 W at 980 nm. The LightForce PRO (LTC-900) is indicated for emitting energy in the infrared spectrum to provide topical heating for the purpose of elevating tissue temperature for temporary relief of minor muscle and joint pain, muscle spasm, pain and stiffness associated with arthritis and promoting. relaxation of the muscle tissue and to temporarily increase local blood circulation. The PRO is a compact, portable medical laser therapy system. The laser light delivery system consists of a flexible optical fiber threaded through a lightweight hand piece. Activation occurs when the operator enables the laser and presses the finger switch. Depending on laser system configuration, the finger switch can function as an on/off switch. A touch-screen display panel allows the operator to adjust or set laser output level. The laser can operate in continuous wave mode or controlled pulse mode.

Recalls

DateFirmReason
2026-01-09Technological Medical Advancements LLCSoftware update is needed for laser systems, for topical infrared heating to provide temporary relief of minor muscle and joint pain, spasms, stiffness associated with minor arthri
2025-07-24THOR Photomedicine LtdThe potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
2025-07-24THOR Photomedicine LtdThe potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
2025-07-24THOR Photomedicine LtdThe potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
2025-07-24THOR Photomedicine LtdThe potential for a failed welded nut to shear off. When this occurs, the gas strut on one side of the device becomes detached from the canopy, making the canopy difficult to lift.
2025-02-12MITO RED LIGHT INCThe near-infrared therapy device contains a lithium-ion battery that may malfunction, causing it to catch fire or burst into flames.
2023-11-20THOR Photomedicine LtdOne of the ball studs attaching a gas strut to the canopy may fail. There is potential risk of injury should a ball stud fail when someone is in the NovoTHOR and trying to open or
2022-09-13Thor Photomedicine LtdFailure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.
2022-09-13Thor Photomedicine LtdFailure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.
2022-09-13Thor Photomedicine LtdFailure of ball stud components causing the gas strut on one side to detach from the canopy, making the canopy difficult to lift.