Home › Hightech American Industrial Labs Inc › EB-2000xyz
EB-2000xyz
Hightech American Industrial Labs Inc · Model 1 · UDI-DI 00852194002021
9
510(k) clearances
0
Adverse events
2
Recalls
10
Facilities
Description
HAI EB-2000xyz is the non-contact eyebank specular microscope which is to be used to evaluate the endothelial cells layer for layer for both human and animal cornea. HAI EB-2000xyz used the same design as the Bio-Optic Inc. (An US manufacturer), it considered the device substantially equivalent to legally marketed predicate devices market predicate devices market in interstate commerce prior to May 28, 1976. That do not require approval of a premarket approval. According to the letter from FDA on April. 26, 2094.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2019-02-25 | Nidek, Inc. | The electromagnetic noise exceeded the upper limit of the Electromagnetic Compatibility Standard (EMC: IEC 60601-1-2) |
| 2016-08-08 | Nidek Inc | Software version 1.08 and 1.09 for the Specular Microscope CEM 530 included a change of analysis results feature that was not reviewed and approved by the FDA. |