FDA Device Intelligence
HomeDECIMAL, INC. › .decimal Proton Range Compensator

.decimal Proton Range Compensator

DECIMAL, INC. · Model 2.8 · UDI-DI 00851147007014

IXI Prescription
1
510(k) clearances
38
Adverse events
6
Recalls
12
Facilities

Description

Custom manufactured solid Range Compensators for intensity modulation of external beam proton radiation therapy.

Recalls

DateFirmReason
2013-08-15Elekta, Inc.Clinical mistreatment may occur when there are errors in the clinical workflow, or settings are incorrect.
2013-05-01Elekta, Inc.It is possible for an intermittent error to occur in the gantry sensor of the DMLC IV Systems
2013-01-18Elekta, Inc.If the Apex collimator is incorrectly installed on the radiation head, a touchguard inhibit activates the view restrictions dialog box only shows the top 12 restrictions.
2012-02-07Siemens Medical Solutions USA, IncThis correction is being performed to update the safety knob of the Siemens brand mMLC (ModuLeaf) to conform to tie IEC 60601-1 safety standard.
2011-02-03Xoft Inc.Xoft Axxent Flexishield Mini product may may shed particles identified as tungsten, and may look like suspicious calcifications on follow up scans if the product was used during In
2008-03-14Philips Medical Systems North America Co. PhillipsPotential for radiologic patient table to become immobile and unable to move again due to force sensor sensitivity to electromagnetic radiation.