Home › ACTUATED MEDICAL, INC. › GripTract-GI Endoscopic Tissue Manipulator

GripTract-GI Endoscopic Tissue Manipulator

ACTUATED MEDICAL, INC. · Model GT-CL130 · UDI-DI 00850458007171

FDF Prescription
1
510(k) clearances
84,519
Adverse events (63 deaths)
25
Recalls
77
Facilities

Description

GripTract-GI Model GT-CL130 is a single-use, non-sterile, endoscope accessory intended to assist in positioning the distal end of an endoscope from the mucosal surface and assist with optical visualization, diagnosis, and endoscopic treatment. GripTract Model GT-CL130 is indicated for use in the large intestine with any standard endoscope with a distal tip outer diameter of 12.8 – 13.3 mm and a working length of 130 -133 cm.

Recalls

DateFirmReason
2024-01-11Olympus Corporation of the AmericasSpecific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in tempera
2024-01-11Olympus Corporation of the AmericasSpecific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in tempera
2024-01-11Olympus Corporation of the AmericasSpecific repaired colonovideoscopes were assembled without an adhesive which protects them from certain physical conditions such as high levels of vibration, fluctuation in tempera
2023-10-16Aizu Olympus Co., Ltd.Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in
2023-10-16Aizu Olympus Co., Ltd.Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in
2023-10-16Aizu Olympus Co., Ltd.Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in
2023-10-16Aizu Olympus Co., Ltd.Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in
2023-10-16Aizu Olympus Co., Ltd.Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in
2023-10-02Olympus Corporation of the AmericasSterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel.
2023-09-25Aizu Olympus Co., Ltd.Channel air drying process was not validated, and that a small percentage of scopes returned to customers after repair had a wet channel. Water remaining in the channel of the endo