FDA Device Intelligence
Home › True Indicating LLC › Larr

Larr

True Indicating LLC · Model CSPN-15L · UDI-DI 00850064244045

JOJ
10
510(k) clearances
392
Adverse events (1 deaths)
24
Recalls
129
Facilities

Description

The Larr Type 5 Integrating Indicator for Steam CSPN-15L is designed to chemically react over time with the critical parameters of steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles: Gravity Displacement: 121 ° C for 30 minutes, 132° C for 15 minutes, 132° C for 25 minutes, 132° C for 3 minutes, 135° C for 10 minutes; Dynamic Air Removal (Vacuum Assist): 132° C for 4 minutes, 135° C for 3 minutes. Minimum Stated Values as determined in a resistometer: SV121 °C/16.5 min., SV132°C/2.0min, SV134 °C/1.4 min. SV135°C/1.2 min.

Recalls

DateFirmReason
2023-12-07TERRAGENE S.A.Due to a discrepancy between the FDA-Cleared Indications for Use (IFU) and the actual IFU distributed with the products.
2023-04-063M Company - Health Care Business3M Attest Steam Chemical Integrators with dates of manufacture between November 6, 2022 and January 17, 2023 have an increased potential to leak ink during a sterilization cycle.
2023-04-063M Company - Health Care Business3M Attest Steam Chemical Integrators with dates of manufacture between November 6, 2022 and January 17, 2023 have an increased potential to leak ink during a sterilization cycle.
2022-08-10Steris Corporation Hopkins FacilityProduct contains incorrect chemical indicators in packaging.
2022-05-25Mckesson Medical-Surgical Inc. Corporate OfficeFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 0
2022-05-25Mckesson Medical-Surgical Inc. Corporate OfficeFacility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 0
2021-12-29Young Dental Mfg Co I LLCThe Dry Heat Indicator Labels packaging may contain TTS Indicator Labels instead of Dry Heat Indicator Labels.
2020-06-17Resource Optimization & Innovation LLCThe firm received reports from customer that the Regard Migrating Indicator shown malfunctions and does not go into the "accept" window at the time of use.
2013-12-19Steris CorporationSTERIS has identified that the outer shipping package for one lot (Lot #024552A) of VERIFY SixCess Challenge Packs incorrectly identifies the lot as Lot #024522A.
2013-06-10Steritec Products Mfc. Co., Inc.Mesa Labs, based out of Bozeman, MT, informed SteriTec Products, Inc. of a product recall involving the EZTest Steam and Smart-Read Biological Indicators (BI) because the readout t

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.