OCCObaby
DEOCCO LLC · Model OCCOflex 43 · UDI-DI 00850016971104
Description
Digital Thermometer enables fast and reliable measurements. These thermometers provide very high clinical accuracy and have been designed to provide maximum user-friendliness. Digital Thermometer series can be used in conjunction with or without a disposable probe cover when preferred. The basic principle of these thermometers is that a change of thermistor resistance, caused by changes in temperature, is converted to changes of frequency of R-C oscillator circuit. Therefore, the temperature can be given by measuring the frequency of the oscillator. For a given time period by applying to R-C oscillator circuit, changes of temperature will correspond to changes of pulse number.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2025-09-17 | Baxter Healthcare Corporation | Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, w |
| 2024-06-13 | Baxter Healthcare Corporation | The Braun Thermoscan PRO 6000 ear thermometers may have been shipped with a compact disc (CD) containing an outdated version of the Instructions for Use (IFU). |
| 2021-07-20 | Welch Allyn Inc | If the device is exposed to fluid ingress and is used before the cleaning fluid has completely dried, there is a risk of the device overheating potentially causing a burn to the us |
| 2019-10-28 | Cardinal Health 200, LLC | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( |
| 2019-10-28 | Cardinal Health 200, LLC | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( |
| 2019-10-28 | Cardinal Health 200, LLC | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( |
| 2019-10-28 | Cardinal Health 200, LLC | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( |
| 2019-10-28 | Cardinal Health 200, LLC | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( |
| 2018-02-02 | Philips Consumer Lifestyle | Device may provide inaccurate measurements. |
| 2017-10-17 | 3M Company - Health Care Business | During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower |