Revogene
MERIDIAN BIOSCIENCE, INC. · Model 610210 · UDI-DI 00840733102318
Description
The system is a clinical multiplex instrument intended to measure and sort multiple signals generated my multiple probes, intercalating dyes, or other ligands in an assay from a clinical sample. Signals may be generated by fluorescence or other phenomena and may be measured using filters on a photodiode or other detector. It may integrate sample and/or reagent handling, amplification, dedicated instrument control, data acquisition software, raw data storage mechanisms and other essential hardware components along with the signal reader unit. The system is used with specific assays to comprise an assay test system. A real-time thermocycler is intended to identify and/or quantify the presence of specific sequences of double stranded DNA, amplified from a biological source and labeled with fluorescently labeled probes or through the use of intercalating dyes and detect using a high-power light-emitting diode (LED). Fluorescence emission is detected through the use of filters on a photodiode. The emission filters are optimized for use with specific fluorescent dyes.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2025-03-20 | Abbott Molecular, Inc. | The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the foot |
| 2024-12-10 | Biocartis Nv | False-positive MSI-H results generated by the Idylla MSI Test. |
| 2024-03-20 | Abbott Molecular, Inc. | The Alinity m System camera detects whether tubes are "capped." If the camera detects a cap and the retention bar is not present on the sample rack, the system will generate the me |
| 2024-02-01 | Abbott Molecular, Inc. | Products that were intended for export only which were not approved or registered in accordance with U.S. specifications were shipped to U.S. customers. |
| 2023-12-15 | Hologic, Inc. | Due to potential false positive Flu B results when a sample was also SARS-CoV-2 positive. |
| 2023-06-14 | DiaSorin Molecular LLC | There is a potential for the direct amplification disc to malfunction which may result in spillage of liquid from the disc. |
| 2022-12-09 | Abbott Molecular, Inc. | The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have |
| 2022-11-11 | Abbott Molecular, Inc. | The instrument was found to be installed without plunger assemblies in the clamp bar for an Amplification Detection Unit. After correction of the issue, the instrument was returne |
| 2022-10-12 | Abbott Molecular, Inc. | Abbott has identified potential performance issues for the Alinity m System software version 1.6.5: 1) In a unique scenario, the waste chute flapper was found to not open when th |
| 2022-06-22 | Meridian Bioscience Inc | When a run is aborted, the cooling period protection does not occur and the user is able to open the lid prior to the completion of the cooling period, exposing the user to heated |