FDA Device Intelligence
Home › CAROLINA LIQUID CHEMISTRIES CORPORATION › CAROLINA LIQUID CHEMISTRIES CORP.® MALB MICROALBUMIN CC662

CAROLINA LIQUID CHEMISTRIES CORP.® MALB MICROALBUMIN CC662

CAROLINA LIQUID CHEMISTRIES CORPORATION · Model CC662 · UDI-DI 00817302020977

DCFJIT Prescription
10
510(k) clearances
64
Adverse events
3
Recalls
40
Facilities

Description

MALB (Microalbumin) 2 x 60 tests

Recalls

DateFirmReason
2019-10-14Ortho Clinical Diagnostics IncPotential for the instability of the diluents VITROS 7% BSA (Product Code 8262487) and VITROS Urine Electrolyte Diluent (UED) (Product Code 1110352) when stored in a 2mL cup for 7
2017-04-24Beckman Coulter Inc.Beckman Coulter has confirmed that MA Reagent Lot M609458 may recover outside the lower limit of the published insert range for Beckman Coulter Urine Protein Control Level 1.
2013-04-30Wako Life Sciences, Inc.Product fails to meet the accuracy at the low end of the measuring range.

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.