Home › KONICA MINOLTA MEDICAL IMAGING U.S.A., INC. › Overhead Tube Crane
Overhead Tube Crane
KONICA MINOLTA MEDICAL IMAGING U.S.A., INC. · Model XR5200 · UDI-DI 00817100020094
KPR
Prescription
10
510(k) clearances
1,233
Adverse events (5 deaths)
168
Recalls
83
Facilities
Description
NOVA CSAP Ceiling suspension tube stand Motorized with Auto positioning and touch screen panel (to work with auto position table and wall stand versions).
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2026-01-09 | FUJIFILM Healthcare Americas Corporation | It was found that the contacts of the electromagnetic contactors used inside the cabinet of the X-ray high-voltage generator were welded, and the current might continue to be suppl |
| 2025-10-01 | Philips Medical Systems DMC GmbH | Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/ph |
| 2025-10-01 | Philips Medical Systems DMC GmbH | Philips has identified that some ceiling suspension telescopic carriages manufactured from July 2024 through June 2025 may be difficult to move vertically (i.e. additional force/ph |
| 2025-09-15 | FUJIFILM Healthcare Americas Corporation | It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the |
| 2025-03-10 | Philips North America Llc | Potential for collimator to fall as a result of incorrect installation. |
| 2025-03-10 | Philips North America Llc | Potential for collimator to fall as a result of incorrect installation. |
| 2025-01-28 | GE Medical Systems, LLC | The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure tech |
| 2025-01-28 | GE Medical Systems, LLC | The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure tech |
| 2025-01-28 | GE Medical Systems, LLC | The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure tech |
| 2025-01-28 | GE Medical Systems, LLC | The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure tech |
Query this across every FDA database — by API.
Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.