Description

Endoscopic Polypectomy Snare, Small Oval - Hot

Recalls

DateFirmReason
2026-08-19Olympus Corporation of the AmericasPotential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
2026-08-19Olympus Corporation of the AmericasPotential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
2026-08-19Olympus Corporation of the AmericasPotential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
2026-08-19Olympus Corporation of the AmericasPotential for detachment of ceramic tip due to manufacturing without the designated amount of adhesive.
2026-01-07Olympus Corporation of the AmericasDevices which did not undergo thermoforming could deform and lose performance.
2026-01-07Olympus Corporation of the AmericasDevices which did not undergo thermoforming could deform and lose performance.
2026-01-07Olympus Corporation of the AmericasDevices which did not undergo thermoforming could deform and lose performance.
2026-01-07Olympus Corporation of the AmericasDevices which did not undergo thermoforming could deform and lose performance.
2026-01-07Olympus Corporation of the AmericasDevices which did not undergo thermoforming could deform and lose performance.
2026-01-07Olympus Corporation of the AmericasDevices which did not undergo thermoforming could deform and lose performance.