FDA Device Intelligence
Home › BIOFIRE DIAGNOSTICS, LLC › BioFire® Blood Culture Identification 2 (BCID2) Panel

BioFire® Blood Culture Identification 2 (BCID2) Panel

BIOFIRE DIAGNOSTICS, LLC · Model RFIT-ASY-0147 · UDI-DI 00815381020338

PAMPEOPEN Prescription
2
510(k) clearances
643
Adverse events (8 deaths)
11
Recalls
6
Facilities

Description

IVD reagent kit containing 30 tests.

Recalls

DateFirmReason
2024-06-20Luminex CorporationPotential for defective test cartridges which may result in false positive results for the Staphylococcus lugdunensis and Enterococcus faecium targets.
2023-11-07Luminex CorporationPotentially defective utility trays in the reagent kits.
2020-10-13Luminex CorporationLuminex has determined that three (3) separate test fixtures used by field service engineers to service the VERIGENE SP Processors were procured from unapproved sources and, in som
2020-06-15Luminex CorporationAn instrument failure mode that may result in a Blood Culture Gram Positive (BC-GP) or Gram Negative (BC-GN) false negative call.
2018-12-26Luminex CorporationOn lot of VERIGENE(R) BC-GP Utility Kits (20-011-018) were labeled with the VERIGENE(R) Amplification Tray label.
2018-09-14Luminex CorporationPotential to lead to a false negative results.
2018-09-05BioFire Diagnostics, LLCIncreased risk of false positive results for Proteus when the product is used with a specific blood culture bottles.
2016-12-12Great Basin Scientific, Inc.The Great Basin Staph ID/R Panel is being recalled because the S. aureus probe could potentially result in a false positive.
2014-05-27BioFire Diagnostics, Inc.BioFire has identified an increased risk of false positive results when the FilmArray Blood Culture Identification (BCID) Panel is used with bioMrieux BacT/ALERT Standard Anaerobic
2013-10-10Nanosphere, Inc.Nanosphere has received several reports relating to an increased rate of Processing Errors during procedures performed on the Processor SP caused by Tips in Tip Holder Assemblies f