FDA Device Intelligence
Home › Zoll Medical Corporation › Powerheart® AED G3 Plus Semi Automatic

Powerheart® AED G3 Plus Semi Automatic

Zoll Medical Corporation · Model 9390E-1001 · UDI-DI 00812394023325

MKJ Prescription
10
510(k) clearances
270,708
Adverse events (4,538 deaths)
336
Recalls
68
Facilities

Description

G3Plus, AED, Semi Automatic, AHA 2010, with adult pads and battery

Recalls

DateFirmReason
2026-07-27HeartSine Technologies LtdProducts were distributed without the required Maintenance Supplement, which describes an extra check every 3 months in addition to the monthly check.
2026-07-27HeartSine Technologies LtdDue to a damaged (over stretched) membrane tail that causes an intermittent fault.
2026-04-13Physio-Control, Inc.Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the i
2026-01-21Physio-Control, Inc.Due to pediatric defibrillator electrode delamination
2025-09-18HeartSine Technologies LtdDue to customer complaints, defibrillator pads may contain damage pins that disrupt the mechanical fit and electrical contact.
2025-07-22Physio-Control, Inc.Due to required inspections not being performed on products/units that have gone through servicing.
2025-07-22Physio-Control, Inc.Due to required inspections not being performed on products/units that have gone through servicing.
2025-07-22Physio-Control, Inc.Due to required inspections not being performed on products/units that have gone through servicing.
2025-06-30HeartSine Technologies LtdDue to a component manufacturing issue, Automated External Defibrillator may not function properly (deliver shocks)
2025-03-18Defibtech, LLCIt was noticed there was no Italian translation included for the AED. Article 16 of the Swiss Medical Devices Ordinance (MedDO) states that product information (labelling and instr

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.