FDA Device Intelligence
Home › CLINICAL DIAGNOSTIC SOLUTIONS, INC. › CDS Hematology Calibrator

CDS Hematology Calibrator

CLINICAL DIAGNOSTIC SOLUTIONS, INC. · Model 1x3.0mL · UDI-DI 00812137010360

KRZ
10
510(k) clearances
21
Adverse events
2
Recalls
18
Facilities

Description

CDS Hematology Calibrator is manufactured for calibration of multi-parameter hematology analyzers for WBC, RBC, Hgb, MCV, Plt and MPV, as applicable to the analyzer system. Multi-parameter hematology analyzers require periodic calibration in order to produce accurate results on patient samples. Calibration may be accomplished by transferring calibration factors to the analyzer through fresh blood samples that have been assayed by reference methods. However, a more direct and convenient approach is to use a calibrator material such as CDS Hematology Calibrator, having system-specific assigned values, derived from a carefully calibrated analyzer or group of analyzers. CDS Hematology Calibrator assigned values are derived from replicate analyses on whole blood calibrated hematology analyzers (see reference procedures). Replicate analyses are performed on the user’s analyzer and calibration factors are computed by comparing recovered values and assigned values. The CDS Hematology Calibrator values are then divided by these determined factors to provide calibration settings for making adjustments to the instrument. Hct, MCH, MCHC, RDW and the WBC differential parameters do not require calibration.

Recalls

DateFirmReason
2025-06-04Siemens Healthcare Diagnostics, Inc.Products have the incorrect platelet (PLT) value assignments for the ADVIA 120/2120/2120i Hematology Systems.
2011-10-25Bio-Rad Laboratories, Inc.It was discovered that the Variant II Turbo Link Calibrator Values card contains incorrect values for IFCC units. The numbers are missing a decimal point.

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.