Home › HORIBA INSTRUMENTS INCORPORATED › Pointe BUN on BS480
Pointe BUN on BS480
HORIBA INSTRUMENTS INCORPORATED · Model 14-B7552-600 · UDI-DI 00811727015747
CDQ
Prescription
1
510(k) clearances
347
Adverse events
6
Recalls
39
Facilities
Description
Pointe BUN packaged for use on the BS480 analyzer. 8x60mL, 8x15mL Kit. For the quantitative determination of urea nitrogen in serum. For in vitro diagnostic use only.
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2023-08-08 | Siemens Healthcare Diagnostics, Inc. | There is a potential for a positive bias on Quality Control (QC) and patient sample results when using some IRON_2 reagent wedge pairs. When the issue occurs, the observed bias wil |
| 2023-06-13 | Randox Laboratories Ltd. | Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if t |
| 2018-03-27 | Siemens Healthcare Diagnostics, Inc. | Urea Nitrogen reagent kit demonstrated an increased incidence of Calibration Failures. Results cannot be generated by the system when calibration fails. |
| 2016-02-16 | Siemens Healthcare Diagnostics, Inc. | Siemens Healthcare Diagnostics has determined that Dimension Vista Blood Urea Nitrogen (BUN) lots 15215AE, 15243BB, 15264BA, 15299BB, 15300BA and 15341AC may exhibit inaccurate pat |
| 2011-10-03 | Abbott Laboratories, Inc | The reagents are being recalled due to particulate matter in some of the reagent cartridges. The particulate matter has been identified as Aureobasidium fungus species. |
| 2003-11-24 | Abbott Laboratories Diagnostic Div | False low results may occur with out error codes due to substrate depletion in samples with more than 220mg/dL BUN. |