Total Bilirubin Reagent, for the quantitative determination of total bilirubin in serum on Hitachi analyzers. For in vitro diagnostic use only. Size: 1 x 2L, 1 x 500ml
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2024-07-30 | Siemens Healthcare Diagnostics, Inc. | During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling |
| 2023-08-05 | Ortho-Clinical Diagnostics, INc. | Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate de |
| 2020-07-13 | Ortho-Clinical Diagnostics | Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides |
| 2020-07-13 | Ortho-Clinical Diagnostics | Etrombopag Interference for VITROS Chemistry Products: TBIL Slides, VITROS BuBc Slides and VITROS XT TBIL-ALKP Slides |
| 2020-02-13 | Siemens Healthcare Diagnostics, Inc. | Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy |
| 2020-02-13 | Siemens Healthcare Diagnostics, Inc. | Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy |
| 2019-09-11 | Roche Diagnostics Operations, Inc. | Due to low Quality Control recovery and invalid Calibration. |
| 2019-08-15 | Beckman Coulter Inc | It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quanti |
| 2019-08-15 | Beckman Coulter Inc | It has been determined that N-acetyl p benzoquinone imine (NAPQI), a metabolite of acetaminophen (paracetamol), may cause negative interference for assays if present in high quanti |
| 2018-12-17 | Roche Diagnostics Corporation | New endogenous interference claims have been established for the following assays used on the cobas c 311 analyzer and the cobas c 501, 502, 701, 702, and Modular Analytics P/D mod |