Home › HORIBA INSTRUMENTS INCORPORATED › Pointe Scientific, Inc.
Pointe Scientific, Inc.
HORIBA INSTRUMENTS INCORPORATED · Model HI904-600 · UDI-DI 00811727012708
JIY
Prescription
10
510(k) clearances
55
Adverse events
19
Recalls
36
Facilities
Description
Iron (Ferrozine) Reagent, for the quantitative determination of iron in serum on Hitachi analyzers. For in vitro diagnostic use only. Size: 5 x 100ml,1 x 100ml
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2025-11-03 | GET TESTED INTERNATIONAL AB | Distribution without premarket approval/clearance. |
| 2024-03-13 | Siemens Healthcare Diagnostics, Inc. | Potential for falsely elevated Chol_2, LDLC, and Trig_2 results on the Atellica CH and Atellica CI analyzers when the previous result in the cuvette was Iron3. Results in a positiv |
| 2023-09-29 | Sentinel CH SpA | A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for |
| 2023-09-29 | Sentinel CH SpA | A high recovery of Iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of Quality Control and patient samples results, up to 15% for |
| 2022-06-28 | Siemens Healthcare Diagnostics, Inc. | Falsely Elevated Atellica CH Microalbumin_2 (ALB_2) Results due to Reagent Carryover from the Iron_2 Assay. This defect occurs only when a microalbumin test is run immediately aft |
| 2016-08-25 | Siemens Healthcare Diagnostics, Inc. | Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instruct |
| 2016-08-25 | Siemens Healthcare Diagnostics, Inc. | Siemens Healthcare Diagnostics has become aware of deferoxamine interference with Dimension and Dimension Vista IRON assays at a different concentration than listed in the Instruct |
| 2016-07-20 | Sterling Diagnostics, Inc. | Sterling Diagnostics, Inc. initiated a recall on 7/20/2016 for multiple diagnostic kits due to a recision of the previously extended expiration dates for the kits. |
| 2013-06-12 | Ortho-Clinical Diagnostics | Ortho Clinical Diagnostics is recalling Slide Dispense cartridges used with some VITROS Chemistry Product Slides due to slide dispense issues. |
| 2012-10-08 | Siemens Healthcare Diagnostics, Inc. | Siemens Healthcare Diagnostics has confirmed the potential for falsely elevated results with IRON Flex(R) reagent cartridges when tests are processed from a well set contaminated w |
Query this across every FDA database — by API.
Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.
Get your free API keyRead the docsData compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.