FDA Device Intelligence
HomeRIVERPOINT MEDICAL, LLC › JuggerKnot

JuggerKnot

RIVERPOINT MEDICAL, LLC · Model CM-99115 · UDI-DI 00810020087147

MBI
10
510(k) clearances
25,622
Adverse events (5 deaths)
113
Recalls
276
Facilities

Description

Soft Anchor

Recalls

DateFirmReason
2025-09-23Biomet, Inc.There is a potential for a missing slotted button assembly in one lot of ZipTight, Acute AC Joint Implant, Single Ziploop
2025-09-09Biomet, Inc.The outer carton box label is for the JuggerKnot 1.0mm Mini products. However, the products within the carton are JuggerKnot 1.4mm Short devices.
2025-01-29Stryker CorporationIncrease in complaints concerning "pullwire" breakage that may result in the broken pullwire remaining in the anchor implant body after deployment.
2024-10-29Smith & Nephew Inc.Sterile barrier breach due to inadequate packaging design that could not hold the geometry of the device.
2023-10-25Riverpoint Medical, LLCUse of a 1.4 drill in hard bone conditions caused increased insertion forces leading to subsequent failures of the inserter, when the inserters bend, it may also move off the drill
2023-08-23Stryker CorporationExpired Product distributed
2023-05-04T.A.G. MEDICAL PRODUCTS CORPORATION, LTD.Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients include
2023-05-04T.A.G. MEDICAL PRODUCTS CORPORATION, LTD.Drill guides might bend when axial forces are applied by surgeon during procedure and result in potential metal shavings due to friction with the drill bit.Risk to patients include
2022-07-13Paragon 28, Inc.Suture anchor may break upon insertion, which could lead to potential debris in patient, and delays in surgery.
2020-11-25Biomet, Inc.Specific lots of multiple sterile medical devices may have a breach in the sterile seal. A breach in the packaging seal could lead to loss of sterility of the device.