CMV IgG Ab - IMMULITE - RGT - 100 Tests

Recalls

DateFirmReason
2025-11-03GET TESTED INTERNATIONAL ABDistribution without premarket approval/clearance.
2022-10-19Bio-Rad Laboratories, Inc.Due to an unusual increase in the positivity rate with human IgM antibodies to cytomegalovirus (CMV) Enzyme Immunoassay (EIA).
2017-01-11BioMerieux SADuring development of the VIDAS 3 software version 1.2, some anomalies have been identified and observed to be already present in the current software version VIDAS 3 version 1.1.4
2015-03-26Diamedix CorporationProduct contained an incorrect substrate.
2009-01-19Abbott LaboratoriesDefective caps: Customers have observed reagent bottles not opening appropriately, which could lead to a probe crash on the instrument.
2007-04-23Abbott LaboratoriesFailure of the AxSYM instrument to open (actuate) some reagent packs from certain lots. The defects can result in probe crashes if not detected prior to placement of the reagent k
2004-06-09bioMerieuxThe product may give false positive results.
2003-06-18Meridian Bioscience IncThe Low Positive Standard kit component is losing potency that may result in false positive test results.
2003-06-12Zeus Scientific IncThe Calibrator included within the kit may be losing reactivity.