Recalls
| Date | Firm | Reason |
|---|---|---|
| 2009-05-07 | Beckman Coulter Inc. | The recall was initiated after Beckman Coulter confirmed reports of inaccurate Access Intrinsic Factor Ab test results displayed on, or printed from, the UniCel Dxl portion of an i |
| 2006-03-08 | Abbott Laboratories MPG | The 30-day onboard storage information is not included in the barcode labels for these two lots. As a result, the ARCHITECT system software is unable to track how long a reagent k |
| 2004-08-19 | Abbott Laboratories HPD/ADD/GPRD | Changes in the ARCHITECT B12 Reagent lots may adversely affect the shape of the calibration curve, over time. These changes may produce controls out of range low and similar decrea |