FDA Device Intelligence
Home › Siemens Healthcare Diagnostics Inc. › RAPIDPOINT®340/350 pCO2 Electrode

RAPIDPOINT®340/350 pCO2 Electrode

Siemens Healthcare Diagnostics Inc. · Model 10375797 · UDI-DI 00630414566252

CHLJGSJFPCEMGKF
1
510(k) clearances
2,419
Adverse events (49 deaths)
185
Recalls
34
Facilities

Description

RAPIDPOINT® 340/350 PCO2 Electrode for use with Rapidpoint® 340/350 Analyzers.

Recalls

DateFirmReason
2026-08-18Radiometer Medical ApSIn vitro diagnostic analyzer had a risk of potentially inaccurate results requiring changes to instructions for use and: 1) Sample types, including whole blood samples no longer su
2026-05-21Instrumentation LaboratoryThe active ingredient concentration in the lysing bag solution of specific GEM Premier 5000 lots did not meet established manufacturing specifications. The malfunction could result
2026-05-21Instrumentation LaboratoryThe active ingredient concentration in the lysing bag solution of specific GEM Premier 7000 with iQM3 PAK lots did not meet established manufacturing specifications. The malfunctio
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-05-08Instrumentation LaboratoryPotential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.
2026-04-02Siemens Healthcare Diagnostics IncUse of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.