Home › SAKURA FINETEK U.S.A., INC. › Tissue-Tek® Cryo3®
Tissue-Tek® Cryo3®
SAKURA FINETEK U.S.A., INC. · Model 5808 · UDI-DI 00615233079756
1
510(k) clearances
40
Adverse events
10
Recalls
21
Facilities
Description
Specimen Holders, Medium
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2025-01-30 | LEICA BIOSYSTEMS NUSSLOCH GMBH | Firm became aware of software issues related to most recent firmware update which could result in loss of patient tissue. |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2024-09-04 | LEICA BIOSYSTEMS NUSSLOCH GMBH | As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potent |
| 2019-05-03 | Leica Microsystems, Inc. | The current labelling does not include any specific warning about usage of flammable freezing sprays. Such flammable freezing sprays can potentially ignite and can cause serious i |
| 2010-09-23 | Thermo Fisher Scientific | Through internal quality review, firm determined the need to conduct a more comprehensive system validation to support the claim of complete bacterial, fungal and viral disinfectio |
| 2010-09-23 | Thermo Fisher Scientific | Through internal quality review, firm determined the need to conduct a more comprehensive system validation to support the claim of complete bacterial, fungal and viral disinfectio |