Geratherm Daisy Color Pacifier Digital Thermometer
R.G. ENTERPRISES (U.S.A.), INC. · Model RETAIL Baby Pacifier Thermometer · UDI-DI 00614801300704
Description
Daisy Infant Pacifier Thermometer is a clinical thermometer intended for the measurement of oral body temperature in infants and young children. The device is waterproof, uses a CR2032 (3 V) battery. Dimensions: 60 mm x 50 mm x 42 mm. Weight: Approx. 23 grams (including battery). Beeper: Approx. 5 secs. sound signal when end temperature reached. Memory: Built-in memory enabling display of the previously measured value. Shut-off: Automatic shut-off in approx. 30 seconds. Backlight Color Coded Visual fever alarm indicators green yellow and red : 36,0 T 37,1 green 37,2 T 37,4 yellow 37,5 T 43,9 red A protective storage cover included
Recalls
| Date | Firm | Reason |
|---|---|---|
| 2025-09-17 | Baxter Healthcare Corporation | Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, w |
| 2024-06-13 | Baxter Healthcare Corporation | The Braun Thermoscan PRO 6000 ear thermometers may have been shipped with a compact disc (CD) containing an outdated version of the Instructions for Use (IFU). |
| 2021-07-20 | Welch Allyn Inc | If the device is exposed to fluid ingress and is used before the cleaning fluid has completely dried, there is a risk of the device overheating potentially causing a burn to the us |
| 2019-10-28 | Cardinal Health 200, LLC | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( |
| 2019-10-28 | Cardinal Health 200, LLC | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( |
| 2019-10-28 | Cardinal Health 200, LLC | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( |
| 2019-10-28 | Cardinal Health 200, LLC | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( |
| 2019-10-28 | Cardinal Health 200, LLC | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( |
| 2018-02-02 | Philips Consumer Lifestyle | Device may provide inaccurate measurements. |
| 2017-10-17 | 3M Company - Health Care Business | During a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower |