FDA Device Intelligence
HomeA.M.G. Médicale Inc › Physio Logic Accuflex Digital Thermometer

Physio Logic Accuflex Digital Thermometer

A.M.G. Médicale Inc · Model 016-628 · UDI-DI 00057565166283

FLL
10
510(k) clearances
1,303
Adverse events (13 deaths)
18
Recalls
479
Facilities

Description

Accuflex® single-patient digital thermometers can help to prevent the risk of cross-contamination, nosocomial infections and make it easy to prevent the spread of contamination among patients. This digital thermometer is easy to use and provides fast readings. Most importantly, Accuflex is reliable and accurate for patients of all ages. Accurate within ± 0.1°C in the 35°C to 42°C range Fast response time: 10 seconds Washable for easy cleaning Gentle-flex tip for patient comfort Auto shut-off feature Memory recall / display of last 30 readings Individually packaged, includes storage case Anti-Bacterial additive Convenient colour codes – blue for oral and red for rectal

Recalls

DateFirmReason
2025-09-17Baxter Healthcare CorporationAffected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, w
2024-06-13Baxter Healthcare CorporationThe Braun Thermoscan PRO 6000 ear thermometers may have been shipped with a compact disc (CD) containing an outdated version of the Instructions for Use (IFU).
2021-07-20Welch Allyn IncIf the device is exposed to fluid ingress and is used before the cleaning fluid has completely dried, there is a risk of the device overheating potentially causing a burn to the us
2019-10-28Cardinal Health 200, LLCThe frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (
2019-10-28Cardinal Health 200, LLCThe frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (
2019-10-28Cardinal Health 200, LLCThe frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (
2019-10-28Cardinal Health 200, LLCThe frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (
2019-10-28Cardinal Health 200, LLCThe frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (
2018-02-02Philips Consumer LifestyleDevice may provide inaccurate measurements.
2017-10-173M Company - Health Care BusinessDuring a recent investigation, 3M confirmed that a programming translation error could occur in a small amount of the sensors that could lead to a temperature readout that is lower