FDA Device Intelligence
Home › ARKRAY USA, INC. › AUTION™ MAX AX-4030 Analyzer

AUTION™ MAX AX-4030 Analyzer

ARKRAY USA, INC. · Model 14505 · UDI-DI 00015482145057

KQO
1
510(k) clearances
1,091
Adverse events (1 deaths)
35
Recalls
68
Facilities

Description

Urine Analyzer utilize reflectance technology with  with AUTION Strips reagent to provide qualitative or semiquantitative urine chemistry results.  Refractometery is used for specific gravity measurements, light transmission for color, and light scatter for turbidity.

Recalls

DateFirmReason
2026-08-12Beckman Coulter, Inc.The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replace
2026-08-12Beckman Coulter, Inc.The rinse-line check valve has a shorter service lifespan than originally expected and requires annual replacement to prevent lamina fluid leakage; however, the check valve replace
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag Previous Sample Had Sperm was enabled but not displayed so that a carryover event could have be
2023-04-05Beckman Coulter, Inc.Beckman Coulter has become aware of an intermittent issue in which the optional flag "Previous Sample Had Sperm" was enabled but not displayed so that a carryover event could have
2021-10-18Roche Diagnostics Operations, Inc.A potential risk for false negative nitrite results exists when endogenous creatinine levels are 15,000 mg/L and above. No interference of Nitrite results was observed at creatinin
2020-03-30Beckman Coulter Inc.A defect in the SATA power adapter cable within the instrument s computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or me
2020-03-30Beckman Coulter Inc.A defect in the SATA power adapter cable within the instruments computer has the potential risk for an electrical short that may lead to the following outcomes: charring and/or mel
2020-02-17Beckman Coulter, Inc.The Leukocyte and Compensation pads were switched on some vials of lot number 7212269M of the iChem VELOCITY Strips.

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Data compiled from public U.S. FDA sources (GUDID, 510(k)/PMA, MAUDE, recalls, establishment registration). Adverse-event reports are unverified and do not establish causation. Informational only — not medical, clinical, or regulatory advice.