Recent recalls

DateFirmReason
2024-07-22EVIDENT SCIENTIFIC INCA broken fuse inside the laser that was caused by a short circuit at PIN 4 of the DOOR CTRL. This short circuit was caused by a badly positioned cable that made
2019-10-02Zeiss, Carl IncWhen the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb