Home › Device types › RBY
X-Ray Fluorescence
FDA product code RBY · Class N · Unknown
0
Registered devices
2
Adverse events
5
Recalls
0
510(k) clearances
Recent recalls
| Date | Firm | Reason |
|---|---|---|
| 2017-07-11 | Olympus Scientific Solutions Americas | Instruments may, under certain specific circumstances listed below, fail to stop when a test is initiated and no sample is present (proximity shut down fails to |
| 2017-01-07 | Olympus Scientific Solutions Americas | It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, th |
| 2016-10-27 | Olympus Scientific Solutions Americas | Possible unintentional x-ray emission after users attempt early termination of the sequence. |
| 2016-06-22 | Olympus Scientific Solutions Americas | Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not preven |
| 2016-01-13 | Olympus Scientific Solutions Americas | The Firm has discovered a Software bug. |
Query this across every FDA database — by API.
Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.
Get your free API keyRead the docsSourced from public U.S. FDA data. Informational only — not medical or regulatory advice. Not affiliated with the FDA.