FDA Device Intelligence
HomeDevice types › RBY

X-Ray Fluorescence

FDA product code RBY · Class N · Unknown

0
Registered devices
2
Adverse events
5
Recalls
0
510(k) clearances

Recent recalls

DateFirmReason
2017-07-11Olympus Scientific Solutions AmericasInstruments may, under certain specific circumstances listed below, fail to stop when a test is initiated and no sample is present (proximity shut down fails to
2017-01-07Olympus Scientific Solutions AmericasIt was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, th
2016-10-27Olympus Scientific Solutions AmericasPossible unintentional x-ray emission after users attempt early termination of the sequence.
2016-06-22Olympus Scientific Solutions AmericasTrigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not preven
2016-01-13Olympus Scientific Solutions AmericasThe Firm has discovered a Software bug.

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs
Sourced from public U.S. FDA data. Informational only — not medical or regulatory advice. Not affiliated with the FDA.