Recent recalls

DateFirmReason
2018-01-02Siemens Healthcare Diagnostics, Inc.Concentrations for the level of biotin that does not interfere is not currently listed in the instruction for use for the Dimension Vista CTNI, Dimension Vista
2017-06-29Siemens Healthcare Diagnostics, Inc.Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine
2016-02-09Siemens Healthcare Diagnostics, Inc.Siemens Healthcare Diagnostics has received customer complaints regarding QC shifts and imprecision after one day when using Dimension SIRO lot EB6064. They hav
2010-05-27Abbott LaboratoriesThe Architect Sirolimus assay lot 80162M100 is exhibiting higher than normal reports of barcode read errors.
2010-04-30Abbott LaboratoriesIMx Sirolimus reagent lot 802873106 has exhibited an increased frequency of calibration errors.
2006-08-11Abbott Laboratories MPGThe sirolimus values may shift higher after storage at 2-8C or after one freeze/thaw cycle of specimens.