FDA Device Intelligence
HomeDevice types › JAI

Couch, Radiation Therapy, Powered

FDA product code JAI · Class 2 · Radiology

349
Registered devices
57
Adverse events
18
Recalls
32
510(k) clearances

Devices with this code

Recent recalls

DateFirmReason
2023-10-06LEONI CIA CABLE SYSTEMSUsers of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the m
2023-10-06LEONI CIA CABLE SYSTEMSUsers of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the m
2018-05-16Elekta, Inc.Upon completion of a 6D workflow, the HexaPOD evo Module may be in a tilted state if it is moved to the position used for 3D (non-iGUIDE) treatments resulting i
2017-11-08Med Tec IncWhen an error message remains displayed and is not cleared in the Protura software with the Elekta pedestal coordinates, moving the pedestal could cause the Pro
2017-09-05Forte Automation Systems IncCommunications between the Patient Positioning System and the accuracy filter can periodically fail with no clear indication to the operator.
2016-03-31Med Tec IncWhen using the recalled product in split shift mode, the non-sub-millimeter positioning has been found to be inconsistent with the setting parameters.
2015-01-29Elekta, Inc.It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed
2014-04-17Elekta, Inc.If screws are not properly tightened with a torque wrench they may become loose over time.
2013-12-05Siemens Medical Solutions USA, IncUpdate to correct the possibility for patients or operators to have their fingers trapped or pinched between the table top and the bottom frame of the 550 TxT T
2013-03-27Accuray IncorporatedA5 gearboxes responsible for pitch (head up and head down) and roll (patient left and right) motions of the RoboCouch manipulator may be defective.

Query this across every FDA database — by API.

Linked GUDID · 510(k)/PMA · MAUDE · recalls · establishments. Free tier, no card required.

Get your free API keyRead the docs